Why Your Function Generator Outputs Twice the Set Voltage
Set 2 Vpp but measure 4 Vpp? Your generator is not broken — it assumes a 50 Ω load. Learn why the voltage doubles on a scope and how to set the load right.

You set the generator to 2 V peak-to-peak, connect it to your oscilloscope, and the scope reads 4 V. The first instinct is to send the generator back — but in most cases, both instruments are fine. The display does not show the voltage the generator produces. It shows the voltage you would get with a 50 Ω load connected. A scope input is about 1 MΩ, almost an open circuit, so nearly all of the internal voltage appears at the load: roughly double the displayed value. Every major generator maker documents this trap. The fix is a menu setting, not a repair.
The 50 Ω Assumption Behind the Display
Behind the output socket sits a voltage source with a resistor in series — almost always 50 Ω. Source and load form a divider, so the voltage at the load is:
Vout = Vsource × Rload / (Rload + 50 Ω)
Generator makers calibrate the display against a matched 50 Ω load. Ask for 2.00 V peak-to-peak and the source behind the resistor is really producing 4.00 V; into a matched load, half is dropped internally and the load receives exactly 2.00 V. Attach anything else and the arithmetic moves. Tektronix works the same example in its AWG support note: set to 2 V peak-to-peak, a 50 Ω load sees 2 V, a 1 kΩ load sees 3.81 V, and a 1 MΩ scope input sees about 4 V.
The 50 Ω value matches the coaxial cables used at radio frequencies and stops reflections from a mismatched far end.
The Load Setting Most People Miss
Many generators let you tell them what they are driving: 50 Ω or high impedance — the one setting most often left at its default. On Keysight's modular generators, switching termination from 50 Ω to high impedance doubles the displayed amplitude — set 5 V peak-to-peak and the panel reads 10 V. The offset doubles the same way: 100 mV set for a 50 Ω load becomes 200 mV into an open circuit. But the setting does not work the same way on every instrument. On some handheld generators the 50 Ω / high-Z switch changes only the display: EEVblog users found this on Siglent's SAG1021i, where flipping the setting changed the screen number while the scope kept reading 5.6 V peak-to-peak. Never assume — check the manual, then measure.
Why Double Voltage Is More Than a Display Error
A factor of two sounds cosmetic until it reaches the circuit: a generator will happily present ten volts to a three-volt logic input.
Overdriving the device under test. Twice the intended drive can exceed a pin's maximum rating, stress a sensor input, or saturate an amplifier.
A doubled offset. The offset is subject to the same divider. A small DC offset you set deliberately can double at a high-impedance input, shifting a bias point or a comparator threshold.
Silent clipping. On many instruments the sum of amplitude and offset is limited, so a large amplitude plus a large offset quietly clips the waveform — the shape on the scope looks wrong, and the cause is a limit you never saw.
At higher frequencies, an unterminated cable also reflects the signal back, adding ringing to fast edges.
A Five-Minute Check
If the scope disagrees with the generator, work through this order first:
- Set a known signal: sine wave, 1 kHz, 1 V peak-to-peak, output on, connected straight to the scope with a BNC cable — no probe, no circuit in between.
- Flip the load setting. Switch between 50 Ω and high impedance and watch both numbers. The display should change; if the scope reading stays the same, the setting is display-only on that model.
- Terminate the line. Add a 50 Ω feed-through terminator at the scope end; the reading should fall to the displayed value. If it does, the generator was right all along.
- Check the probe. A 10:1 probe with the scope still set to 1:1 reads one tenth of the real voltage — a 1 V peak-to-peak sine reads 120 mV. Match the probe switch to the channel setting.
- Check the cable, then the calibration. If the reading is still wrong with a proper termination, swap the cable and check the generator against a second meter. Only now is a fault likely.
What to Check Before You Order a Batch
For a distributor or own-brand buyer, this trap is a specification question: a generator that is "wrong" in a customer's lab becomes a return, a complaint and a lost reorder. Check these points before committing to a batch:
Read the amplitude specification properly. Does the datasheet state amplitude into 50 Ω, into an open circuit, or both? A figure quoted without a load is not a specification — ask the supplier to state it in writing.
Check whether the load setting exists. Some generators have an adjustable load menu; many compact and handheld models do not. Handheld 3-in-1 instruments such as the HDS242 — a 40 MHz oscilloscope, multimeter and waveform generator in one unit, with a DDS generator covering 10 Hz to 156 kHz — and the ZT-703S, a 50 MHz dual-channel scope with a built-in signal generator, save bench space for field service, but a built-in generator is often fixed at 50 Ω with no load menu. Confirm the output specification in the manual for the exact model.
Test a sample both ways. During sample evaluation, set 2 V peak-to-peak and measure it twice: once into a 50 Ω terminator, once into a scope input. With the load setting correct, both readings should agree with the display — one test that catches the problem before it reaches your customers.
Count the total cost. A 50 Ω feed-through terminator costs very little; a prototype board damaged by double drive, or a batch returned because "the amplitude is wrong", costs far more.
Ask your supplier these five questions before you order:
- What is the generator's output impedance, and is it fixed or switchable?
- Is the amplitude specified into 50 Ω, into high impedance, or both?
- Does the load setting change the actual output, or only the displayed value?
- What is the maximum amplitude into an open circuit?
- Can we test one sample with a 50 Ω terminator before the batch ships?
The Bottom Line
A generator that reads double on the scope is almost never broken. It is a 50 Ω source driving a 1 MΩ load, and the display was calibrated for a load that is not there. Set the load to match what you are actually driving, confirm it with a terminator, and the "fault" disappears. The amplitude number means nothing until you know the load it was measured into.
If your bench team keeps seeing double, check the load setting before you return the instrument. For handheld oscilloscopes with a built-in waveform generator, or a standalone generator for your product line, contact the team at zydhardware.com — tell us the load your circuit presents.